Elemental Instruments, a division of Ohio Lumex logo

Measuring Pericellular Oxygen Tension In Cell Culture Using PyroScience Oxygen Sensors

Facebook
Twitter
LinkedIn

Oxygen tension plays a key role in organic tissue function. When studied in controlled environments, it can reveal useful insights about human physiology and pathophysiology.

Mammalian cell culture research is a commonly used proxy method for studying conditions in the human body. Last year, a group of scientists from the Northeastern University in Boston, MA came up with a new method of measuring oxygen tension in oxygen controlled mammalian cell culture environments.

Fig. 1: Measuring pericellular oxygen (O2) tension in mammalian cell culture using sensor spots. (A) Self-adhesive sensor spots (OXSP5-ADH) were placed on the surface on which cells were grown. 6-well plate wells are shown as an example. (B) Cable adapters (SPADBAS) were glued (SPGLUE) onto the backside of the cell culture vessel. Cables (SPFIB-BARE) were fastened within the adapters and connected to a meter (FSO2-C4).

Having assumed that the gas phase oxygen concentration equals the pericellular oxygen concentration, they used special oxygen controlled incubators along with the cell culture wells where PyroScience sensor spots were installed (see Fig. 1). In their experiment, they examined the relationship between gas phase and pericellular oxygen tensions in standard physioxic (5% O2) and hypoxic (1% O2) cultures.

Fig. 2: Cellular O2 consumption can induce hypoxia and anoxia in physioxic (5% O2) and hypoxic (1% O2) cell culture, respectively. (A-B) Pericellular O2 profiles (left) and average O2 tension values (right) of media, human mammary epithelial, MCF7, or MDA-MB-231 in a 24-well plate in a 5% O2 (A) or 1% O2 (B) incubator for 48h.

As a result of this study, the researchers discovered that pericellular oxygen tensions are far below incubator setpoints in both standard physioxic and hypixoc cell culture protocols. They concluded that in order to accurately control for oxygen in cell cultures, it is highly recommended to measure pericellular oxygen tension.

For more information about this research along with its methods and conclusions, see here.

Facebook
Twitter
LinkedIn
Ohio Lumex logo in white

Search

Request a Parts Quote

Onsite Services Quote

Request a Rental

Request a Quote